Journal of Nuclear Fuel Cycle and Waste Technology 2015;13(2):123-130. Published online: Jun, 30, 2015
DOI : dx.doi.org/10.7733/jnfcwt.2015.13.2.123
For the analysis of 90Sr, which is a pure beta emitter, radiochemical separation from the main interfering elements such as Ca, Ba and Ra is required due to their similarity in chemical behavior to strontium. This study describes a radioanalytical procedure using extraction chromatography for separating Sr from interfering elements. The maximum capacity of the resin for Sr was approximately 6 mg per 1.5 mL of bed volume (BV). The recovery of Sr on the resin 1.5 mL (BV) was quantitative for the calcium level of 200 mg at the flow rate of 1 mL min-1. However the chemical yield declined by increasing the flow rate by up to 5 mL min-1 even at the calcium level of 200 mg. When using the same BV of Sr resin, the performance of the resin was enhanced as the cross-sectional area of the Sr resin column is small.
Keywords
Radiostrontium,Sr resin,Automated radionuclide separator,Optimization