Journal of Nuclear Fuel Cycle and Waste Technology 2006;4(3):275-284. Published online: Sep, 30, 2006
In the measurement of gamma rays emitted from the nuclide in the radioactive waste drum, to analyze the nuclide concentration accurately, it is necessary to use the proper calibration standards and to correct for the attenuation of the gamma rays. Two drums having a different density were used to analyze the nuclide concentration inside the drum in this study. After carrying out the system calibration, we measured the gamma rays emitted from the standard source inside the model drum with changing the distance between the drum and the detector. The measured values were corrected with the three kinds of gamma attenuation correction methode, as a results, the error was less than 10 % in the low density drum and less than 25 % in the high density drum. The measured activity in the short distance was more accruable than in the long distance. The transmission correction for the mass attenuation showed good results(very low error) compared to the mean density and the differential peak correction method.
Keywords
NDA,Model drum,Gamma-ray,Attenuation correction,Mass attenuation coefficient,Transmission Source