Review

  • Journal of Nuclear Fuel Cycle and Waste Technology
  • Volume 6(2); 2008
  • Article

Journal of Nuclear Fuel Cycle and Waste Technology 2008;6(2):111-117. Published online: Jun, 30, 2008

PDF

Phoswich Detector for Simultaneous Measuring Alpha/beta Particles

  • Gye Hong Kim ; Chan Hee Park ; Kune Woo Lee ; Chong Hun Jung ; Bum Kyoung Seo
Abstract

The new type phoswich detector consisting of the ZnS(Ag) and plastic scintillator for alpha/beta-ray simultaneous counting was developed for monitoring radiological contamination inside pipes. The detection performance was estimated using the PSD (pulse shape discrimination) method as a function of distance between the scintillator and radioactive source. The attenuation of particles traveling through a thin film for preventing the detector from being contaminated was experimentally estimated. It is concluded from our investigation that the phoswich detector developed can provide a sufficient alpha/beta-ray discrimination. The application of a thin film for preventing the detector from being contaminated was proven to be feasible.

Keywords